MALDI vs Electrospray: How Peptides Are Ionized for Mass Spectrometry, and the TFA Suppression Problem

How soft ionization made peptide mass spectrometry possible: MALDI matrices, electrospray multiply charged ions, charge-state deconvolution, nanoelectrospray and why TFA suppresses ESI signal.

Share

Every mass spectrum of a peptide starts with the same problem: a mass spectrometer can only measure ions in the gas phase, and a peptide is a large, polar, thermally fragile molecule that does not want to be either gaseous or charged. For decades that problem put peptides and proteins effectively out of reach. The two techniques that solved it — matrix-assisted laser desorption/ionization (MALDI) and electrospray ionization (ESI) — are the reason modern peptide identity testing exists. This article explains how each one works, when a laboratory chooses one over the other, and why a trace of trifluoroacetic acid in the mobile phase can quietly destroy an ESI signal. It is written as analytical background for laboratory researchers.

Why "soft" ionization was the breakthrough

Older ionization methods added so much energy to the molecule that large analytes simply fragmented into unidentifiable debris before a molecular ion could be recorded. "Soft" ionization means transferring an intact molecule into the gas phase as an ion with minimal internal energy, so that the molecular mass survives the trip.

Two independent routes to that goal appeared at the end of the 1980s.

MALDI: a laser and a matrix

In MALDI the analyte is co-crystallised with a large excess of a small organic matrix compound that absorbs strongly at the laser wavelength, typically in the ultraviolet. A short laser pulse hits the crystal, the matrix absorbs the energy and ablates, and the analyte is carried into the gas phase in the expanding plume, picking up a charge — usually a single proton — in the process.

Karas and Hillenkamp reported laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons (Analytical Chemistry, 1988), and in the same year Tanaka and colleagues reported protein and polymer analyses up to m/z 100,000 by laser ionization time-of-flight mass spectrometry (Rapid Communications in Mass Spectrometry, 1988), using a different approach to energy transfer. Together these papers established that intact large biomolecules could be measured.

The matrix is not an inert filler; it is the active component, and matrix choice is part of method development:

  • Sinapinic acid and other cinnamic acid derivatives. Beavis, Chait and Fales described cinnamic acid derivatives as matrices for ultraviolet laser desorption mass spectrometry of proteins (Rapid Communications in Mass Spectrometry, 1989). This family is the traditional choice for larger proteins.
  • α-Cyano-4-hydroxycinnamic acid (CHCA). Beavis, Chaudhary and Chait introduced CHCA as a MALDI matrix (Organic Mass Spectrometry, 1992). It became the workhorse for peptides and tryptic digests, giving fine crystals and good sensitivity in the lower mass range.
  • 2,5-Dihydroxybenzoic acid (DHB). Strupat, Karas and Hillenkamp described DHB as a matrix for laser desorption-ionization mass spectrometry (International Journal of Mass Spectrometry and Ion Processes, 1991). DHB is more tolerant of salts and buffers and is often chosen for glycopeptides and dirtier samples.

MALDI spectra are usually dominated by singly charged ions, which makes them easy to read: one peptide, one peak, at roughly its molecular mass plus one proton. MALDI is also comparatively tolerant of contaminants, and because the sample sits on a plate rather than flowing through the source, an interesting spot can be re-analysed later.

Electrospray: charge from a charged droplet

ESI takes the opposite approach. The sample is in solution, pumped through a narrow capillary held at high voltage. The field disperses the liquid into a fine spray of charged droplets; as solvent evaporates, the droplets shrink until charge repulsion exceeds surface tension and they divide, repeating until bare gas-phase analyte ions remain.

Fenn, Mann, Meng, Wong and Whitehouse reported electrospray ionization for mass spectrometry of large biomolecules (Science, 1989), and the defining feature they described is multiple charging. A peptide or protein picks up several protons, producing a series of peaks at [M+nH]n+ for a range of n. Because the instrument measures mass-to-charge ratio, multiple charging divides the apparent m/z by n, which is what brings large molecules inside the range of ordinary analysers.

The cost is that the spectrum is no longer a simple picture of one molecule as one peak. Mann, Meng and Fenn set out how to interpret mass spectra of multiply charged ions (Analytical Chemistry, 1989): adjacent peaks in a charge-state envelope differ by one proton, so the spacing itself reveals n, and the whole series can be collapsed algorithmically to a single neutral mass. That procedure — deconvolution — is now automatic in every ESI data system, and it is also a quiet quality check: a clean, self-consistent charge-state envelope that deconvolutes to one mass is good evidence that one species is present.

The other decisive advantage of ESI is that it starts from a liquid, so it couples directly to liquid chromatography. LC-MS of peptides is ESI's home ground.

Nanoelectrospray

Reducing the flow rate improves electrospray efficiency rather than hurting it. Wilm and Mann described the analytical properties of the nanoelectrospray ion source (Analytical Chemistry, 1996), operating at flow rates in the nanolitre-per-minute range from a fine pulled capillary. Smaller initial droplets desolvate more readily, ion yield per molecule rises sharply, and a few microlitres of sample can be sprayed for a long time. Nanospray, coupled to nano-flow LC, is the basis of sensitive peptide mapping and proteomics work.

The TFA problem

Trifluoroacetic acid is the default ion-pairing additive in reversed-phase peptide chromatography. It gives sharp peaks and excellent resolution for exactly the reason it causes trouble downstream: it pairs tightly with the basic sites on a peptide that electrospray needs to protonate, and it makes the droplet surface unfavourable for releasing those ions. The result is strong signal suppression — the chromatography improves and the mass spectrometric response collapses, sometimes by more than an order of magnitude.

Two 1995 papers from the same group characterised the effect and described a practical remedy. Kuhlmann, Apffel, Fischer, Goldberg and Goodley reported signal enhancement for gradient reversed-phase HPLC-ESI-MS analysis with trifluoroacetic and other strong acid modifiers by postcolumn addition of propionic acid and isopropanol (Journal of the American Society for Mass Spectrometry, 1995), and Apffel and colleagues reported enhanced sensitivity for peptide mapping with electrospray LC-MS in the presence of signal suppression due to TFA-containing mobile phases (Journal of Chromatography A, 1995). The approach, widely known as the "TFA fix", adds a weaker acid and an organic solvent after the column, displacing the ion pairing and improving desolvation without sacrificing the separation.

The alternatives in routine use are to substitute formic acid for TFA and accept slightly broader peaks, or to keep TFA and apply postcolumn addition. Either way, a method that runs TFA straight into an ESI source and reports low sensitivity is describing a known artefact, not necessarily a poor sample. Note that this is a mobile-phase ionization issue and is separate from TFA present as a counter-ion on the peptide salt itself.

Choosing between them

MALDI is fast, tolerant of salts and buffers, and gives simple singly charged spectra, which suits rapid identity confirmation and intact-mass checks. ESI integrates with chromatography, quantifies well, produces the multiply charged precursors that tandem MS fragmentation methods prefer, and scales down to nanospray sensitivity. Most analytical laboratories treat them as complementary rather than competing, and a thorough characterisation package may include both.

References

  • Karas M, Hillenkamp F. Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons. Analytical Chemistry. 1988;60(20):2299-2301. doi:10.1021/ac00171a028.
  • Tanaka K, Waki H, Ido Y, Akita S, Yoshida Y, Yoshida T, et al. Protein and polymer analyses up to m/z 100,000 by laser ionization time-of-flight mass spectrometry. Rapid Communications in Mass Spectrometry. 1988;2(8):151-153. doi:10.1002/rcm.1290020802.
  • Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246(4926):64-71. doi:10.1126/science.2675315.
  • Mann M, Meng CK, Fenn JB. Interpreting mass spectra of multiply charged ions. Analytical Chemistry. 1989;61(15):1702-1708. doi:10.1021/ac00190a023.
  • Beavis RC, Chait BT, Fales HM. Cinnamic acid derivatives as matrices for ultraviolet laser desorption mass spectrometry of proteins. Rapid Communications in Mass Spectrometry. 1989;3(12):432-435. doi:10.1002/rcm.1290031207.
  • Beavis RC, Chaudhary T, Chait BT. α-Cyano-4-hydroxycinnamic acid as a matrix for matrix-assisted laser desorption mass spectrometry. Organic Mass Spectrometry. 1992;27(2):156-158. doi:10.1002/oms.1210270217.
  • Strupat K, Karas M, Hillenkamp F. 2,5-Dihydroxybenzoic acid: a new matrix for laser desorption-ionization mass spectrometry. International Journal of Mass Spectrometry and Ion Processes. 1991;111:89-102. doi:10.1016/0168-1176(91)85050-V.
  • Wilm M, Mann M. Analytical properties of the nanoelectrospray ion source. Analytical Chemistry. 1996;68(1):1-8. doi:10.1021/ac9509519.
  • Kuhlmann FE, Apffel A, Fischer SM, Goldberg G, Goodley PC. Signal enhancement for gradient reverse-phase high-performance liquid chromatography-electrospray ionization mass spectrometry analysis with trifluoroacetic and other strong acid modifiers by postcolumn addition of propionic acid and isopropanol. Journal of the American Society for Mass Spectrometry. 1995;6(12):1221-1225. doi:10.1016/1044-0305(95)00571-4.
  • Apffel A, Fischer S, Goldberg G, Goodley PC, Kuhlmann FE. Enhanced sensitivity for peptide mapping with electrospray liquid chromatography-mass spectrometry in the presence of signal suppression due to trifluoroacetic acid-containing mobile phases. Journal of Chromatography A. 1995;712(1):177-190. doi:10.1016/0021-9673(95)00175-m.

Research use only. All products sold by Morpheus Research are intended strictly for laboratory research and are not for human or veterinary consumption. Nothing in this article is medical advice, and no statement here is intended to diagnose, treat, cure or prevent any condition. Study findings are reported as published and should not be read as evidence of effects in people.